A Compact oxygen concentrator for premature infants: Integrating portability and performance in the third-generation prototype

Authors

  • Cornelia Christiani Vianney, S.T. Departemen Teknik Mesin, Universitas Indonesia Author
  • Prof. Dr. Ir. Raldi Artono Koestoer Departemen Teknik Mesin, Universitas Indonesia Author
  • Prof. Dr. Yudan Whulanza, S.T., M.Sc. Departemen Teknik Mesin, Universitas Indonesia Author
  • Prof. Dr. Ir. Imansyah Ibnu Hakim, M.Eng Departemen Teknik Mesin, Universitas Indonesia Author

DOI:

https://doi.org/10.71452/zs2cbg95

Keywords:

oxygen concentrator, portable, infant, design development

Abstract

Oxygen concentrators are vital devices in medical oxygen therapy. It provides supplemental oxygen, extracted from the environmental air, to patients suffering from hypoxemia or other clinical conditions characterized by low blood oxygen saturation. The need for oxygen therapy extends across various patient populations, including adults, children, and neonates, especially premature infants. However, commercially available concentrators are usually designed for adults, making them unsuitable for premature infants whose delicate, underdeveloped organs can be harmed by high oxygen levels. Meanwhile, field observations from the Indonesian Incubator Team indicate frequent demand for supplemental oxygen following hospital discharge. To meet the urgent need for a suitable device, especially for premature infants discharged from hospitals, this study describes the design and development of a compact, portable, and neonatal-specific oxygen concentrator. A reverse engineering approach was adopted through a three-phase prototyping process. The first prototype verified the feasibility of the pressure swing adsorption (PSA) principle as the main process of the oxygen concentrator. The second prototype integrated a complete system to evaluate performance by measuring flow rate and oxygen concentration and adjusting components to achieve optimal results. The final, third prototype refined the design to improve compactness and portability. The device achieved an oxygen concentration of about 40% oxygen purity with a steady flow rate of 2 L/min. This developed oxygen concentrator presents a significant step forward in neonatal care, providing a tailored and accessible solution that can support widespread use in home-based infant healthcare programs.

References

R. R. Vemula, M. Kothare, and S. Sircar, "Novel Design and Performance of a Medical Oxygen Concentrator Using a Rapid Pressure Swing Adsorption Concept," AIChE Journal, vol. 60, 2014, doi: 10.1002/aic.14518.

S. Ongtrakul, A. Thitiratannapong, C. Pintavirooj, and T. Treebupachatsakul, "Pressure Swing Absorption Oxygen Concentrator equipped with Remote Monitoring Pulse Oximeter," in 2021 13th Biomedical Engineering International Conference (BMEiCON), ed, 2021, pp. 1-5.

W. Stoudemire, "Common respiratory disorders in children," in Encyclopedia of Child and Adolescent Health, First Edition, vol. 1: Elsevier, 2023, pp. 600-616.

F. Jabeen, R. Ahmad, N. Khursheed, S. Rashid, A. Ali, and S. A. Ganie, "Pharmacogenomics of pulmonary and respiratory diseases," in Pharmacogenomics: from Discovery to Clinical Implementation: Elsevier, 2023, pp. 223-246.

T. Keim, R. Amjad, and R. Fales, "Modeling and feedback control of inspired oxygen for premature infants," in ASME 2011 Dynamic Systems and Control Conference and Bath/ASME Symposium on Fluid Power and Motion Control, DSCC 2011, 2011, vol. 1, pp. 501-508, doi: 10.1115/DSCC2011-6107. [Online]. Available: https://www.scopus.com/inward/record.uri?eid=2-s2.0-84881413229&doi=10.1115%2fDSCC2011-6107&partnerID=40&md5=aafa912898eb223f5b9b02bb2e094adc

W. Tin, "Optimal oxygen saturation for preterm babies: Do we really know?," in Biology of the Neonate, 2004, vol. 85, 4 ed., pp. 319-325, doi: 10.1159/000078173. [Online]. Available: https://www.scopus.com/inward/record.uri?eid=2-s2.0-3042677619&doi=10.1159%2f000078173&partnerID=40&md5=da91dd53142adb1aacd6818047a6a6d7

C. C. Vianney, A. Riantono, and R. A. Koestoer, "Design and Prototyping Oxygen Concentrator Simulator toward Low-cost and Portable System," in The 3rd of International Mechanical and Industrial Engineering Conference (IMIEC) 2022, ed. Universitas Negeri Malang, Indonesia, 2022.

F. F. Amarullah, C. C. Vianney, A. Riantono, I. Roihan, and R. A. Koestoer, "Design and Prototyping Low-Cost Portable Oxygen Concentrator for Infant," in International Symposium on Biomedical Engineering (ISBE 2023), ed. Bali, Indonesia, 2023.

K. Johnson, S. D. Scott, and K. D. Fraser, "Oxygen use for preterm infants: Factors that may influence clinical decisions surrounding oxygen titration," Advances in Neonatal Care, Review vol. 11, no. 1, pp. 8-14, 2011, doi: 10.1097/ANC.0b013e318206d0c0.

P. Chanyagorn and P. Kiratiwudhikul, "Automation control algorithms in gas mixture for preterm infant oxygen therapy," in 2015 IEEE International Conference on Consumer Electronics - Taiwan, ICCE-TW 2015, 2015, pp. 236-237, doi: 10.1109/ICCE-TW.2015.7216874. [Online]. Available: https://www.scopus.com/inward/record.uri?eid=2-s2.0-84959537034&doi=10.1109%2fICCE-TW.2015.7216874&partnerID=40&md5=e1b5f9c1e667af274c33e4d9e62c3aeb

P. Kiratiwudhikul and P. Chanyagorn, "Gas mixture control system for oxygen therapy in pre-term infants," in Proceedings of 2014 International Conference on Smart Computing, SMARTCOMP 2014, 2014, pp. 289-294, doi: 10.1109/SMARTCOMP.2014.7043870. [Online]. Available: https://www.scopus.com/inward/record.uri?eid=2-s2.0-84946531367&doi=10.1109%2fSMARTCOMP.2014.7043870&partnerID=40&md5=8a3eca30a50151044f9554d7f3cb7cdc

L. Torrejón-Rodríguez, Á. Solaz-García, I. Lara-Cantón, A. Pinilla-González, M. Aguar, and M. Vento, "Clinical Parameters in the First 5 Minutes after Birth Have a Predictive Value for Survival of Extremely Preterm Infants," Maternal-Fetal Medicine, Article vol. 5, no. 4, pp. 244-247, 2023, doi: 10.1097/FM9.0000000000000206.

N. Finer and T. Leone, "Oxygen saturation monitoring for the preterm infant: The evidence basis for current practice," Pediatric Research, Review vol. 65, no. 4, pp. 375-380, 2009, doi: 10.1203/PDR.0b013e318199386a.

A. Buckmaster, G. Arnolda, I. Wright, and J. Foster, "Targeted oxygen therapy in special care nurseries: Is uniformity a good thing?," Journal of Paediatrics and Child Health, Article vol. 48, no. 6, pp. 476-482, 2012, doi: 10.1111/j.1440-1754.2011.02220.x.

Published

09-03-2026

Conference Proceedings Volume

Section

Design and Advanced Mechanics

How to Cite

[1]
C. C. Vianney, R. A. Koestoer, Y. Whulanza, and I. I. Hakim , Trans., “A Compact oxygen concentrator for premature infants: Integrating portability and performance in the third-generation prototype”, Seminar Nasional Tahunan - Teknik Mesin , vol. 23, no. 1, pp. 91–96, Mar. 2026, doi: 10.71452/zs2cbg95.